Ultrasonic echography (B-Scan)

Product No: P5160300

Principle
The fundamental principles concerning the generation of ultrasonic B-scan images (brightness representation of the reflection amplitudes) are demonstrated with the aid of a simple test ob-ject. The experiment is executed with an ultrasonic echoscope in the pulse-echo-mode and the object is scanned manually. Then, the image quality and the most important image defects will be assessed.

Tasks

  1. Measure the 3 edge lengths of the test block with a vernier calliper and determine the time of flight of the sound for the various edge lengths of the test block with the aid of the measurement software.
  2. Calculate the sound velocity of the test block material and switch the measurement software (A-scan mode) to depth measurement.
  3. Produce two B-scans, one with the 1 MHz probe and one with the 2 MHz probe, with the aid of the measurement software (B-scan mode).
  4. Measure the depth and width of the echoes in the resulting images.
  5. Characterise the quality of the images in terms of their resolution and defects.

Related topics

  • Sound velocity
  • reflection coefficient
  • ultrasonic echography
  • A-scan
  • B-scan
  • greyscale display
  • resolution
  • zone of focus
  • and image artefacts.



Materiallist (Excerpt) Product Amount
Basic Set Ultrasonic echoscope 13921-99 1
Vernier caliper 03010-00 1
Ruler, plastic, 200 mm 09937-01 1

Documentation language: ENG

Literature for this article als follows

Title Product Language
TESS expert Handbuch UltraschallanwendungAnwendungen von Ultraschall in Medizin, Material- 01232-02 ENG

back to corresponding listview

Product No: P5160300

Recommend this page  ∇

X-ray expert unit

This product is a world first

PHYWE presents the new x-ray unit: innovative technology, top safety, maximum comfort!

More