Layer thickness determination

Product No: P1306100

Principle

The method of determining layer thickness with beta back scattering is particularly suitable for double-layer systems in which the two components differ sufficiently in their atomic numbers. For example, the lacquer coating on metals can be controlled and corrected during production. Layers of thickness up to half the radius of beta particles in the material can be determined. With thicker material, the back-scattered electrons, which have already suffered a loss of energy in the scattering process, are completely absorbed in the surface layer.

Task

To use beta radiation to determine the thickness of a sheet of paper placed on a lead plate.




Materiallist (Excerpt) Product Amount
Geiger-Müller-Counter 13606-99 1
Geiger-Mueller Counter tube, type B 09005-00 1
Base plate for radioactivity 09200-00 1
Plate holder on fixing magnet 09203-00 1
Absorption material f.student exp 09014-03 1
Counter tube holder on fix.magn. 09201-00 1
Source holder on fixing magnet 09202-00 1

Literature for this article als follows

Title Product Language
TESS advanced Physik Handbuch Radioaktivität 01155-01 DEU
TESS advanced Physics manual Radioactivity 01155-02 ENG

back to corresponding listview