Helium neon laser, advanced set

Product No: P2260705

Principle

The difference between spontaneous and stimulated emission of light is demonstrated. The beam propagation within the resonator cavity of a He- Ne laser and its divergence are determined, its stability criterion is checked and the relative output power of the laser is measured as a function of the tube's position inside the resonator and of the tube current. The following items can be realized with advanced set 08656.02. By means of a birefringent tuner and a Littrow prism different wavelengths can be selected and quantitatively determined if a monochromator is available. Finally you can demonstrate the existence of longitudinal modes and the gain profile of the He-Ne laser provided an analysing Fabry Perot system is at your disposal.

Tasks

  1. Set up the He-Ne laser. Adjust the resonator mirrors by use of the pilotlaser. (left mirror: VIS, HR, plane; right mirror: VIS, HR, R = 700 mm)
  2. Check on the stability condition of a hemispherical resonator.
  3. Measure the integral relative output power as a function of the laser tube's position within the hemispherical resonator.
  4. Measure the beam diameter within the hemispherical resonator right and left of the laser tube.
  5. Determine the divergence of the laser beam.
  6. Measure the integral relative output power as a function of the tube current.

The He-Ne laser can be tuned using a BFT or a LTP. Longitudinal modes can be observed by use of a Fabry Perot Etalon of low finesse. Remark: These points can only be covered quantitatively if a monochromator and an analysing Fabry Perot system are available.

What you can learn about

  • Spontaneous and stimulated light emission
  • Inversion
  • Collision of second type
  • Gas discharge tube
  • Resonator cavity
  • Transverse and longitudinal resonator modes
  • Birefringence
  • Brewster angle
  • Littrow prism
  • Fabry Perot Etalon



Materiallist (Excerpt) Product Amount
Exp.Set-Helium-Neon Laser 08656-93 1
Lyot-plate with holder and rider 08656-10 1
Littrow-prism with x/y-holder 08656-20 1
Fabry-Perot etalon in x/y-holder 08656-30 1
Sliding device, horizontal 08713-00 1
Protection glasses HeNe-laser 08581-10 1
Photoelement f. opt. base plt. 08734-00 1
Cleaning set for laser 08582-00 1
Diffraction grating, 600 lines/mm 08546-00 1
DMM, auto range, NiCr-Ni thermocouple 07123-00 1
Connection box 06030-23 1
Barrel base PHYWE 02006-55 1
Plate holder 02062-00 1
Slide mount f.opt.profile-bench 08286-00 1
Danger sign -laser- 06542-00 1
Vernier caliper 03010-00 1
Scale, l = 750 mm, on rod 02200-00 1
Resistor 10 Ohm 2%, 2W, G1 06056-10 1
Resistor 100 Ohm 2%, 1W, G1 06057-10 1
Resistor 1kOhm, 1W, G1 39104-19 1
resistor 10 kOhm, 1W, G1 39104-30 1
Resistor 100 kOhm, 1W, G1 39104-41 1
Screen, white, 150x150mm 09826-00 1
Measuring tape, l = 2 m 09936-00 1
Connecting cord, 32 A, 750 mm, red 07362-01 1
Connecting cord, 32 A, 750 mm, blue 07362-04 1
Laboratory Experiments Physics, Chemistry, Biology andApplied Sciences, CD-ROM, incl. operating manuals 16502-42 1

Literature for this article als follows

Title Product Language
TESS expert Handbook Laboratory Experiments Physics 16502-32 ENG

back to corresponding listview