Gas chromatograph, 2-column

Product No: 36657-93

Funciton and Applications

For the separation of undecomposed vaporisable substances up to 250 °C.

Benefits:

  •  fixed in the furnace chamber of the gas chromatograph at the injection and detector block
  • signals can be registered via the recorder output with a tY recorder or a computer

Equipment and technical data:

  • 2 exchangeable separating columns
  • Length: 1 m
  • Diameter: 5 mm
  • Column I polar mobile solvent (polyethylene glycol adipate)
  • column II anon-polar mobile solvent (apiezone fatL).
  • separate gasinlet (via a coupler hose connection ) and gas outlet
  • thermal conductivity detector, which is switched on according to the separating column used.
  • Heating output approx. 180Watt.
  • Temperature adjustment 20...250°C,digital display.
  • Temperature regulation with protection against over heating detector: proportional
  • 2 thermal conductivity measuring cells in bridge connection.

Zero point adjustable

measuring bridge electricity supply: 140 mA DC

Mains connection 230 V~ (other voltages available on request).